Omron Healthcare BP652 Bedienerhandbuch

Stöbern Sie online oder laden Sie Bedienerhandbuch nach Blutdruck-Geräte Omron Healthcare BP652 herunter. Omron Healthcare BP652 blood pressure unit Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken

Inhaltsverzeichnis

Seite 1 - Pressure Monitor

ESPAÑOLENGLISH7 seriesAutomatic Wrist Blood Pressure MonitorModel BP652INSTRUCTION MANUAL

Seite 2 - TABLE OF CONTENTS

10This is a general guideline because blood pressure depends on age and morbidity. A lower home blood pressure goal may be recommended for certain pa

Seite 3 - SAVE THESE INSTRUCTIONS

11BATTERY INSTALLATION CAUTIONUseonly1.5V“AAA”alkalinebatterieswiththisdevice.1. Press the hook on the bottom of the battery cover and lift

Seite 4 - SAFETY INFORMATION

12BATTERY REPLACEMENT Low Battery IndicatorWhen the Low Battery Indicator appears on the display, turn the monitor off and remove all the batteries.

Seite 5

13SETTING THE DATE AND TIMESetthemonitortothecurrentdateandtimebeforetakingameasurementforthersttime. 1. TO START THE SETTINGPress t

Seite 6 - BEFORE TAKING A MEASUREMENT

144. SETTING THE DAYPress the Date/Time Setting button ( ) to advance by increments of one day.Press the Date/Time Setting button ( ) to set the curr

Seite 7 - KNOW YOUR UNIT

15SETTING THE ADVANCED POSITIONING SENSOR (APS)You can change the options for various settings on your monitor. Pressing the SET button to change a se

Seite 8 - UNIT DISPLAY

16SETTING FOR RIGHT OR LEFT WRIST MEASUREMENT1. While in the setting mode, press the SET button until the wrist symbol ( or ) appears on the display

Seite 9

17IfthePositioningindicatoris“on1”or“on2”:on1: The buzzer sounds only when the monitor is in proper position.on2: The buzzer sounds a series

Seite 10 - DISPLAY SYMBOLS

18 CAUTIONRead all of the information in the instruction manual and any otherliterature in the box before operating the unit. CAUTIONThis device is

Seite 11 - BATTERY INSTALLATION

195. Fold the remaining part of the wrist cuff back out of the way.APPLYING THE CUFF ON THE RIGHT WRISTWhen taking a measurement using the right wris

Seite 12

2TABLE OF CONTENTSBefore using the Monitor Introduction ...

Seite 13 - SETTING THE DATE AND TIME

20The monitor is designed to take measurements and store the measurement values in memory for two people using USER ID A and USER ID B. The monitor ca

Seite 14

21TAKING A MEASUREMENT5. Press the START/STOP button.All symbols appear on the display.Thecuffstartstoinateautomatically.Asthecuffinates,t

Seite 15

22USING THE MEMORY FUNCTIONEvery time you complete a measurement the monitor stores the blood pressure and pulse rate for each user (A and B) in the m

Seite 16

23USING THE MEMORY FUNCTIONNOTES:•Thedateandtimearealternatelydisplayedwiththemeasurementvalues.•PresstheMEMbuttonrepeatedlytodisp

Seite 17

24CARE AND MAINTENANCETo keep your digital blood pressure monitor in the best condition and protect the unit from damage, follow the directions listed

Seite 18 - APPLYING THE WRIST CUFF

25ERROR INDICATORSSYMBOL CAUSE CORRECTIONMonitor could not detect pulse wave.Take another measurement and remain still until the measurement is comple

Seite 19

26TROUBLESHOOTING TIPSPROBLEM CAUSES AND SOLUTIONSNo power.No display appears on the unit.Replace both batteries with new ones at the same time. Check

Seite 20 - TAKING A MEASUREMENT

27FCC STATEMENTNote:POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only)This product has been tested and found to comply with the limits for

Seite 21

28Your BP652 IntelliSense® Automatic Wrist Blood Pressure Monitor, excluding the wrist cuff and batteries, is warranted to be free from defects in mat

Seite 22 - USING THE MEMORY FUNCTION

29SPECIFICATIONSModel:BP652 (HEM-6052-Z)Display:LCD Digital DisplayMeasurement Range:Pressure: 0 to 299 mmHg, Pulse: 40 to 180/min.Accuracy/Calibratio

Seite 23

3INTRODUCTIONThank you for purchasing the OMRON® BP652 IntelliSense® Automatic Wrist Blood Pressure Monitor.The monitor is compact and easy-to-use at

Seite 24 - CARE AND MAINTENANCE

Omron Healthcare, Inc.1200 Lakeside DriveBannockburn, IL 60015www.omronhealthcare.comCopyright © 2010 Omron Healthcare, Inc.Made in China 5329415-9A

Seite 25 - ERROR INDICATORS

4SAFETY INFORMATIONTo assure the correct use of the product, basic safety measures should always be followed including the warnings and cautions liste

Seite 26 - TROUBLESHOOTING TIPS

5SAFETY INFORMATION Do not use a cellular phone near the device. It may result in an operational failure. Use only Omron authorized parts and accessor

Seite 27 - FCC STATEMENT

6BEFORE TAKING A MEASUREMENTTo ensure a reliable reading follow these recommendations:1. Avoid eating, drinking alcohol, smoking, exercising, and bat

Seite 28 - LIMITED WARRANTY

7KNOW YOUR UNITMain Unit:Components:2 “AAA” Alkaline Batteries Instruction ManualQuick Start GuideDate/Time setting ( ) buttonDisplayWrist cuff

Seite 29 - SPECIFICATIONS

8UNIT DISPLAYPositioning indicatorDiastolic blood pressureIrregular heartbeat symbolMemory symbol(Displayed when viewing values stored in memory)Avera

Seite 30

9DISPLAY SYMBOLSIRREGULAR HEARTBEAT SYMBOL ( )When the monitor detects an irregular rhythm two or more times during the measurement, the Irregular Hea

Kommentare zu diesen Handbüchern

Keine Kommentare